Keysight N9038A MXE Service Guide 89
Instrument Messages
Event Messages
–221 Settings conflict;
Must apply Amplitude
Correction to make
this unit available
These special units only apply when you are doing antenna measurements,
so you must have a correction that includes Antenna Units enabled
–221 Settings conflict; No
meas frequencies are
above 3.6 GHz
Optimize Preselector can only be performed on frequencies in high band,
that is, above 3.6 GHz. The current setup does not have input frequencies (IF)
in this range so an Optimize Preselector cannot be performed
–221 Settings conflict; no
source selected
You must select a source using Select Source before you can do this
–221 Settings conflict;
Normalize is not
available when Scale
Type = Lin
Normalize does not support Linear amplitude scale, since the results are
always presented as a dB ratio
–221 Settings conflict;
Normalize is not
available while Demod
View is on
The normalization (correction) function cannot be used if you are using the
Demod View
–221 Settings conflict;
Normalize is not
available while Trace
Math is on
The Normalize function works by doing trace manipulation, so if trace math is
on you cannot turn on normalization
–221 Settings conflict;
Only active Antenna
Unit available; no
other Y axis units
When a correction with antenna units is turned on, the only Y-axis units you
can have are those that match the Antenna Unit. Turn off the Correction or
the Antenna Unit under Input/Output, Corrections
–221 Settings conflict;
Option not available
You have attempted to perform an action for which a required option is not
installed
–221 Settings conflict;
Param only available
when DUT is a freq
converter
SCPI only message. The sideband and freq context parameters are only
available when a freq conversion setup is in use. Change setup to contain a
freq conversion to use these parameters
–221 Settings conflict;
Param only available
when External LO Mode
is Swept
SCPI only message. This parameter is only available when the LO mode is set
to Swept. Change the LO Mode to Swept
–221 Settings conflict;
Param only available
when External LO Mode
is Fixed
SCPI only message. This parameter is only available when the LO mode is set
to Fixed. Change the Freq Mode to Fixed
–221 Settings conflict;
Param only available
when Frequency Mode
is Fixed
SCPI only message. This parameter is only available when the Freq mode is
set to Fixed. Change the Freq Mode to Fixed