90 Keysight N9038A MXE Service Guide
Instrument Messages
Event Messages
–221 Settings conflict;
Param only available
when Frequency Mode
is Swept
SCPI only message. This parameter is only available when the Freq mode is
set to Swept. Change the Freq Mode to Swept
–221 Settings conflict;
Param only available
when valid cal data
exists
SCPI only message. The ‘Apply Calibration’ parameter is only available when
the stored cal data matches the current setup. Perform a fresh ‘Calibrate
Now’ or change setup such that current cal data is valid
–221 Settings conflict;
Preamp gain is not
available in this
Mode
Preamp gain correction is not available in some Modes or Measurements
–221 Settings conflict;
Preamp unavailable
with electronic
attenuator on
The electronic attenuator is on. Internal preamp cannot be used while you are
using the electronic attenuator
–221 Settings conflict;
Reference marker must
be in same window
A delta marker and its reference must be in the same window. This error
occurs when you try to turn on a delta marker for which the reference is in a
different window
–221 Settings conflict;
Relative Trigger
needs hardware
support for this meas
To do Relative Triggering in this measurement requires optional hardware
that is not present in this analyzer
–221 Settings conflict;
Scale Type = Lin is
not available when
Normalize is on
Only the Log amplitude scale is available in Normalize, since the results are
always presented as a dB ratio
–221 Settings conflict;
Settings conflict;
Pre-trigger is
insufficient for
demod. Decrease Trig
Delay.
–221 Settings conflict;
Signal Track is not
available with
Continuous Peak
The signal tracking feature cannot be used while you are also using the
continuous peak function
–221 Settings conflict;
Signal Track is only
available in Swept SA
measurement
The signal track functionality can be used when making a swept SA
measurement. It is not available in the SA measurement when you are using
FFT sweeps
Err# Message Verbose/Correction Information