Keysight N9038A MXE Service Guide 93
Instrument Messages
Event Messages
–221 Settings
conflict;Auto Scan
Time/Meas Time do not
apply in Stepped Scan
Type.
The Auto Scan Time/Meas Time are not available when Scan Type =Stepped
Scan
–221 Settings
conflict;Auto Scan
Time/Meas Time do not
apply in Stepped Scan
Type.
The Auto Scan Time/Meas Time are not available when Scan Type =Stepped
Scan
–221 Settings
conflict;ClearList &
Start not available
when ScanSeq =
(Re)measure
Cannot perform Clear List & Start during (Re)measure, because we require
the frequency information of the peaks in signal list to perform (Re)measure
–221 Settings conflict;EMI
Detectors and Average
detector can’t be
used together
You cannot turn on any EMI Detector and Average detector together. They
are always mutually exclusive
–221 Settings
conflict;Freq > 1 GHz
is not available
while RF Input 2
enabled
Frequency is limited to 1GHz while RF Input 2 is enabled
–221 Settings
conflict;Function not
available while
measurement is
running
Settings change is not allowed while measurement is running. You must stop
the current measurement if you wish to change the settings
–221 Settings
conflict;function
unavailable with this
EMC Standard
–221 Settings conflict;Log
Percent Auto Step
Rule does not apply
in Swept Scan Type.
The Log Percent rule is not available when Scan Type=Swept Scan because
we are always doing linear sweep
–221 Settings conflict;QPD
+ EMI Average + RMS
Average is not
allowed
You cannot turn on all 3 EMI detectors together. You must turn off one of the
EMI Detectors before you turn this on
Err# Message Verbose/Correction Information