EasyManuals Logo

Mindray ePM 12MC User Manual

Mindray ePM 12MC
364 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #104 background imageLoading...
Page #104 background image
10 - 16
10.7.5.2 Setting Arrhythmia Alarm Shielding Period
The arrhythmia algorithm can disable alarm light and alarm tone for designated period of time when certain
arrhythmia alarms are detected.
This function is password protected. For more information, see Arrhy Shield Time in 27.4.6The Other Tab.
NOTE
• For the following alarms, alarm light and alarm tone cannot be disabled: HR high, HR low,
Tachycardia, Bradycardia, Afib End, Irr. Rhythm End.
• Alarm indication rules for alarms in the atrial fibrillation chain are the same with those for the
medium priority chains.
• The arrhythmia shielding period is only applicable to the alarms in the medium priority chains and
atrial fibrillation chain. For the alarms in the high priority chain, alarm tone and alarm light are
presented as soon as the alarm condition is detected.
Asystole
VͲFib/VͲTach
VͲTach / Vent Brady
Extreme Tachy/
Extreme Brady
Pause
Nonsus VͲTach /
Vent Rhythm
Tachy/Brady
Run PVCs
Couplet
R on T
Bigeminy
Trigeminy
Multiform PVC
PVC
Pacer not Capture /
Pacer not Pacing /
Missed Beats
HR High /
HR Low
Pauses/min
AͲFib/
AͲFib End
Irr. Rhythm/
Irr.Rhythm End
PVCs/min
PVC Alarm Chain
Beat Detection
Alarm Chain
Rate Alarm Chain AͲFib Alarm Chain
High Priority Medium Priority

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray ePM 12MC and is the answer not in the manual?

Mindray ePM 12MC Specifications

General IconGeneral
BrandMindray
ModelePM 12MC
CategoryMedical Equipment
LanguageEnglish

Related product manuals