27 - 6
27.4.5 The Nurse Call Tab
• The call help function works only when the monitor is connected to the network.
• The call help sound may disturb patients in the same department.
27.4.6 The Other Tab
Menu Item Default Setting Description
Signal Type Continuous • Pulse: the nurse call signal is a pulse signal 
and each pulse lasts one second. When 
multiple alarms simultaneously occur, only 
one pulse signal is outputted. If an alarm 
occurs but the previous one is not cleared, a 
new pulse signal will also be outputted.
•
Continuous: the nurse call signal lasts until 
the alarm ends. That is to say the duration of a 
nurse call signal is equal to that of the alarm 
condition.
Contact Type Normally Open Selects the work mode of the nurse call relay
Alarm Priority High Only Selects the priority of alarms sent to the nurse call 
system
Alarm Type Physiological Only Selects the type of alarms sent to the nurse call 
system.
Receive Call Help On Receives the calling signal if a monitor in the 
same department calls for help.
Section Menu Item Default Setting Description
Alarm Priority ECG Lead Off Low Selects the priority of the ECG lead off alarm.
SpO2 Sensor Off Low Selects the alarm level for SpO
2
 sensor off alarm.
IBP No Sensor Med Selects the alarm level for IBP No Sensor alarm.
CMS/eGW 
Disconnected
Low Selects the priority of the CMS and eGateway 
disconnection alarm.
Alarm Delay Alarm Delay 6 sec •
1 sec ~15 sec: for continuously measured 
parameters, an alarm is not presented if the 
alarm condition is resolved within the 
designated delay time.
•
Off: an alarm is always presented.
The setting of 
Alarm Delay is not applied to the 
apnea alarms and the ST alarms.
ST Alarm Delay 30 sec The monitor does not present the ST alarm if the 
alarm condition is resolved within the delay time.