C - 7
C.1.4 Temperature Default Settings
C.1.5 NIBP Default Settings
Item Default Setting
TXX
(XX refers to
temperature site)
Alarm switch On
High limit 38.0 °C
Low limit 35.0 °C
Priority Med
Alarm Outputs Off
T Alarm switch On
High limit 2.0 °C
Priority Med
Alarm Outputs Off
Item Default Setting
NIBP-S Alarm switch On
High limit Adult: 160 mmHg
Pediatric: 120 mmHg
Neonate: 90 mmHg
Low limit Adult: 90 mmHg
Pediatric: 70 mmHg
Neonate: 40 mmHg
Priority Med
Alarm Outputs Off
NIBP-D Alarm switch On
High limit Adult: 90 mmHg
Pediatric: 70 mmHg
Neonate: 60 mmHg
Low limit Adult: 50 mmHg
Pediatric: 40 mmHg
Neonate: 20 mmHg
Priority Med
Alarm Outputs Off
NIBP-M Alarm switch On
High limit Adult: 110 mmHg
Pediatric: 90 mmHg
Neonate: 70 mmHg
Low limit Adult: 60 mmHg
Pediatric: 50 mmHg
Neonate: 25 mmHg
Priority Med
Alarm Outputs Off