27 - 13
27.9 The Unit Settings
27.10 The Time Settings
27.10.1 The Time Synchronization Tab
27.10.2 The Daylight Savings Time Tab
27.11 The Other Settings
Menu Item Default Setting Description
Height Unit cm Selects measurement unit for each parameter.
Weight Unit kg
ST Unit mV
CVP Unit cmH2O
ICP Unit mmHg
CO2 Unit mmHg
O2 Unit %
Temp Unit °C
Pressure Unit mmHg
Section Menu Item Default Setting Description
Nighttime From 22:00 Selects the nighttime for heart rate statistics.
To 06:00
/Start NTP Time SyncOff
On: enables synchronizing the monitor time with
the NTP server time.
Interval 1 hr Select the time interval for synchronizing the
monitor time with the NTP server time.
Time Server Address / The domain name of the time server.
Time Server / The IP address of the time server.
Network Test / Tests whether the NTP server is properly
connected.
Menu Item Default Setting Description
Auto Daylight Savings Time Off On: auto starts the daylight saving time.
Menu Item Default Setting Description
Barometric Pressure 760 mmHg For the mainstream CO2 module, enter the value
of barometric pressure to which the patient
monitor is exposed to.
Be sure to set the barometric pressure properly.
Improper settings will result in erroneous
measurements.
Notch Frequency 50 Hz Selects notch filter frequency according to the
power line frequency of your country.
Mouse Sensitivity 5 /