14 - 2
14.5 Changing Temperature Settings 
14.5.1 Setting the Temperature Alarm Properties
To set the temperature alarm properties, follow this procedure:
1. Select the temperature numeric area to enter the Temp menu.
2. Select the Alarm tab.
3. Enter the password if required. 
4. Set the alarm properties。
14.5.2 Selecting the Temperature Label
Select the temperature label according to the measurement site. To do so, follow this procedure:
1. Select the temperature numeric area to enter the Temp menu.
2. Select the Setup tab.
3. Set the temperature label.
14.5.3 Displaying the Temperature Difference
To display the temperature difference between two measurement sites monitored by the same temperature 
module, switch on corresponding T. To do so, follow this procedure:
1. Select the temperature numeric area to enter the Temp menu.
2. Select the Setup tab.
3. Switch on ∆T.
14.6 Temperature Troubleshooting
This section lists the problems that might occur. If you encounter the problems when using the equipment or 
accessories, check the table below before requesting for services. If the problem persists, contact your service 
personnel.
• For the physiological and technical alarm messages, see DAlarm Messages.
Problem Solution
Do not see Temp numeric area on the main 
screen
1.  Check that if the Temp parameter switch is enabled. If not, enable 
the Temp measurement. For more information, see 
3.11.1Switching 
On or Off a Parameter
.
2. Check that the connections of the temperature probe and the 
temperature cable are tight.
Measurement fails/’--’ is displayed in the Temp 
numeric area
1. If you are using a disposable probe, check the connection between 
the probe and the temperature cable.
2. Try using a known good probe in case the sensor is damaged.