A - 15
A.13.5 Temp Specifications
A.13.6 NIBP Specifications
Standard Meet the standard of ISO 80601-2-56
Technique Thermal resistance
Operating mode Direct mode
Measurement range 0 to 50 °C (32 to 122 °F)
Resolution 0.1°C
Accuracy ±0.1 °C or ±0.2 °F (excluding probe error)
Refreshing rate ≤1 s
Minimum time for accurate measurement Body surface: <100 s
Body cavity: <80 s
Recovery time <15 s (after defibrillation)
Alarm limit Range Step
TXX High (XX refers to temperature site) (low limit +1.0) to 50.0 °C
(low limit +2.0) to 122.0 °F
0.1 °C
0.1 °F
TXX Low (XX refers to temperature site) 0.1 to (high limit - 1.0) °C
32.2 to (high limit - 2.0) °F
ΔT High 0.1 to 50.0 °C
0.2 to 90.0 °F
Standard Meet standard of IEC 80601-2-30
Technique Oscillometry
Mode of operation Manual, Auto, STAT, Sequence
Auto mode repetition intervals 1, 2, 2.5, 3, 5, 10, 15, 20, 30, 60, 90, 120, 180, 240 or 480 min
STAT mode cycle time 5 min
Max measurement time Adult, Pediatric: 180 s
Neonate: 90 s
Heart rate range 30 to 300 bpm
Measurement ranges
(mmHg)
Adult Pediatric Neonate
Systolic: 25 to 290 25 to 240 25 to 140
Diastolic: 10 to 250 10 to 200 10 to 115
Mean: 15 to 260 15 to 215 15 to 125
Accuracy Max mean error: ±5 mmHg
Max standard deviation: 8 mmHg
Resolution 1mmHg
Initial cuff inflation pressure range
(mmHg)
Adult: 80 to 280
Pediatric: 80 to 210
Neonate: 60 to 140
Default initial cuff inflation pressure
(mmHg)
Adult: 160
Pediatric: 140
Neonate: 90
Software overpressure protection Adult: 297±3 mmHg
Pediatric: 297±3 mmHg
Neonate: 147±3 mmHg