A - 16
Static pressure measurement range 0 mmHg to 300 mmHg
Static pressure measurement accuracy ±3 mmHg
Recovery time <15 s (after defibrillation)
PR
Measurement range 30 to300 bpm
Resolution 1 bpm
Accuracy ±3bpm or ±3%, whichever is greater
Alarm limit Range (mmHg)
Step (mmHg)
NIBP-S High Adult: (low limit + 5) to 285
Pediatric: (low limit + 5) to 235
Neonate: (low limit + 5) to 135
NIBP ≤ 50: 1
NIBP > 50: 5
NIBP-S Low 26 to (high limit - 5)
NIBP-M High Adult: (low limit + 5) to 255
Pediatric: (low limit + 5) to 210
Neonate: (low limit + 5) to 120
NIBP-M Low 16 to (high limit - 5)
NIBP-D High Adult: (low limit + 5) to 245
Pediatric: (low limit + 5) to 195
Neonate: (low limit + 5) to 110
NIBP-D Low 11 to (high limit - 5)
NIBP-S Extreme High NIBP-S high limit < 50
Adult: (NIBP-S high limit + 1) to 290
Pediatric: (NIBP-S high limit + 1) to 240
Neonate: (NIBP-S high limit + 1) to 140
NIBP-S high limit ≥ 50
Adult: (NIBP-S high limit + 5) to 290
Pediatric: (NIBP-S high limit + 5) to 240
Neonate: (NIBP-S high limit + 5) to 140
NIBP ≤ 50: 1
NIBP > 50: 5
NIBP-S Extreme Low NIBP-S low limit ≤ 50
25 to (NIBP-S low limit - 1)
NIBP-S low limit > 50
25 to (NIBP-S low limit - 5)
NIBP-M Extreme High NIBP-M high limit < 50
Adult: (NIBP-M high limit + 1) to 260
Pediatric: (NIBP-M high limit + 1) to 215
Neonate: (NIBP-M high limit + 1) to 125
NIBP-M high limit ≥ 50
Adult: (NIBP-M high limit + 5) to 260
Pediatric: (NIBP-M high limit + 5) to 215
Neonate: (NIBP-M high limit + 5) to 125
NIBP-M Extreme Low NIBP-M low limit ≤ 50
15 to (NIBP-M low limit - 1)
NIBP-M low limit > 50
15 to (NIBP-M low limit - 5)