C - 16
C.1.10 BIS Default Settings
C.2 Routine Default Settings
C.2.1 Alarm Default Settings
Scale O
2
: 400 mmHg
CO
2
: 50 mmHg
N
2
O: 50%
Hal, Enf, and Iso: 2.5%
Sev: 4.0%
AA and Des: 9.0%
Waveform Type Draw (for CO
2
only)
O2 Compensation OR: 100%
Other departments: Off
Item Default Setting
Item Default Setting
BIS Alarm switch On
High limit 70
Low limit 20
Priority Med
Alarm Outputs Off
Smoothing Rate 15 sec
Display OR: BIS Trend
Other departments: EEG LT/RT
Scale 100 v
Speed 25mm/s
Filter On
Auto Check On
Select Parameter SQI, EMG, SR, SEF
BIS Expand View
Default Setting
EEG Waveforms All
Parameter 1 BIS L
Parameter 2 EMG
Trend Length 60 min
Item Default Setting
Alarm Volume 2
High Alarm Volume Alarm Volume + 3
Reminder Volume 2
Apnea Delay Adult: 20 sec
Pediatric: 20 sec
Neonate: 15 sec