27 - 5
27.4.3 The Latching Tab
27.4.4 The Remote View Tab
Menu Item Default Setting Description
Lethal  Visible Unselected Selects alarm latching rules: 
•If 
Visible is selected, you can separately latch 
visual alarm signal.
• Latching audible alarm signal simultaneously 
latches visual signal.
• Selecting alarms of lower priority 
simultaneously latches higher priority alarms.
Audible
High Visible
Audible
Med Visible
Audible
Low Visible
Audible
Menu Item Default Setting Description
Reset Remote Bed’s Alarms Off Selects whether you can reset alarms occurring 
to the remote devices from your monitor.
On: the Alarm Reset button appears on the 
bottom left of the 
Remote View screen.
Alarm Reset By Other Bed On
On: alarms on your monitor can be reset by 
remote devices.
Alarm Reminder Visible+Audible Selects what alarm indicators are necessary for 
the remote devices.
•
Visible + Audible: the monitor provides 
visual alarm indication, and continuous 
audible alarm indication if the alarm persists 
at the remote device.
•
Visible + Single Tone: the monitor provides 
visual alarm indication, and a single tone 
when the alarm occurs at the remote device.
•
Visible Only: the monitor only provides visual 
alarm indication.
Alarm Priority All Selects what priority of remote device alarms are 
presented for audible notification
•
All: the monitor sounds if an alarm occurs.
•
High & Med: the monitor sounds if a high or 
medium priority alarm occurs.
•
High Only: the monitor sounds only if a high 
priority alarm occurs.
Alarm Sound ISO Selects the alarm tone pattern for the remote 
device alarms.
Remote Disconnected Alarm On Selects whether an alarm is issued if a remote 
device is disconnected.