A - 22
Mainstream CO
2 
Module
A.13.10 AG Specifications
Standard Meet the standard of ISO 80601-2-55
CO
2
 Measurement range 0 to 150 mmHg
Accuracy 0 to 40 mmHg: 
41 to 70 mmHg: 
71 to 100 mmHg: 
101 to 150 mmHg: 
±2 mmHg
±5% of the reading
±8% of the reading
±10% of the reading
Accuracy drift Meet the requirement for measurement accuracy within 6 hours
Resolution 1 mmHg
Recovery time <15 s (after defibrillation)
Rise time <60 ms
Data sample rate 100 Hz
awRR measurement range 0 to 150 rpm
awRR measurement accuracy ±1 rpm
awRR resolution 1 rpm
Standard Meet the standard of ISO 80601-2-55
Technique Infrared absorption, paramagnetic properties for O
2
 monitoring
Recovery time <15 s (after defibrillation)
Warm-up time Iso accuracy mode: 
Full accuracy mode: 
45 s
10 min
Sample flow rate Adult, Pediatric:
Neonate:
Accuracy: 
 200 ml/min
120 ml/min
±10 ml/min or ±10%, whichever is 
greater
Measurement range CO
2
: 
O
2
: 
N
2
O: 
Des: 
Sev: 
Enf: 
Iso: 
Hal: 
awRR: 
0 to 30%
0 to 100%
0 to 100%
0 to 30%
0 to 30%
0 to 30%
0 to 30%
0 to 30%
2 to 100 rpm
Resolution CO
2
: 
O
2
: 
N
2
O:
Des: 
Sev: 
Enf: 
Iso: 
Hal: 
awRR: 
0.1%
1%
1%
0.1%
0.1%
0.1%
0.1%
0.1%
1 rpm
Iso accuracy  As full accuracy specifications, but derated as follows:
Add ±0.3%
ABS
 to accuracy for CO
2
Add ±8%
REL
 to accuracy for all anesthetic gases
N
2
O accuracy is ±(8%
REL
+2%
ABS
)