Full accuracy Gases Range (%REL)1 Accuracy (%ABS)
CO
2
0≤CO
2
≤1
1<CO
2
≤5
5<CO
2
≤7
7<CO
2
≤10
CO
2
>10
±0.1
±0.2
±0.3
±0.5
Not specified
N
2
O0 - 20
20 - 100 (excluding 20)
±2
±3
O
2
0 - 25
25 - 80 (excluding 25)
80 - 100 (excluding 80)
±1
±2
±3
Des 0 - 1
1 - 5 (excluding 1)
5 - 10 (excluding 5)
10 - 15 (excluding 10)
15 - 18 (excluding 15)
>18
±0.15
±0.2
±0.4
±0.6
±1
Not specified
Sev 0 - 1
1 - 5 (excluding 1)
5 - 8 (excluding 5)
>8
±0.15
±0.2
±0.4
Not specified
Enf, Iso, Hal 0 - 1
1 - 5 (excluding 1)
>5
±0.15
±0.2
Not specified
awRR 2 to 60 rpm
>60 rpm
±1 rpm
Not specified
Note
1
: The highest GAS LEVEL for a single halogenated anaesthetic gas in a gas mixture that is concealed when the
anaesthetic concentration falls is 0.15/0.3% (Full/ISO accuracy).
Accuracy drift Meet the requirement for measurement accuracy within 6 hours
Apnea alarm time 10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s
Refreshing rate ≤1 s
Rise time
(10% ~ 90%)
Gas sample flow rate 120ml/min, using a DRYLINE II neonatal watertrap and
sampling line (2.5m):
CO
2
N
2
O
Hal, Iso, Sev, Des
Enf
O
2
≤250 ms
≤250 ms
≤300 ms
≤350 ms
≤600 ms
Gas sample flow rate 200ml/min, using the adult DRYLINE II water trap and
sampling line (2.5m):
CO
2
N
2
O
O
2
Hal, Iso, Sev, Des
Enf
≤250 ms
≤250 ms
≤500 ms
≤300 ms
≤350 ms
Delay time <4 s