C - 14
C.1.9 Gas Default Settings
Item Default Setting
EtCO2 Alarm switch On
High limit Adult, Pediatric: 50 mmHg
Neonate: 45 mmHg
Low limit Adult, Pediatric: 25mmHg
Neonate: 30mmHg
Priority Med
Alarm Outputs Off
FiCO2 Alarm switch On
High limit 4 mmHg
Priority Med
Alarm Outputs Off
EtO2 Alarm switch On
High limit
88%
Low limit 18%
Priority Med
Alarm Outputs Off
FiO2 Alarm switch On
High limit Adult, Pediatric: 100%
Neonate: 90%
Low limit 18%
Priority Med
Alarm Outputs Off
EtN2O Alarm switch On
High limit 55%
Low limit 0%
Priority Med
Alarm Outputs Off
FiN2O Alarm switch On
High limit 53%
Low limit 0%
Priority Med
Alarm Outputs Off
EtAA/FiAA Alarm switch On
High limit 30%
Low limit 0.0%
Priority Med
Alarm Outputs Off