EasyManuals Logo

Mindray ePM 12MC User Manual

Mindray ePM 12MC
364 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #324 background imageLoading...
Page #324 background image
C - 14
C.1.9 Gas Default Settings
Item Default Setting
EtCO2 Alarm switch On
High limit Adult, Pediatric: 50 mmHg
Neonate: 45 mmHg
Low limit Adult, Pediatric: 25mmHg
Neonate: 30mmHg
Priority Med
Alarm Outputs Off
FiCO2 Alarm switch On
High limit 4 mmHg
Priority Med
Alarm Outputs Off
EtO2 Alarm switch On
High limit
88%
Low limit 18%
Priority Med
Alarm Outputs Off
FiO2 Alarm switch On
High limit Adult, Pediatric: 100%
Neonate: 90%
Low limit 18%
Priority Med
Alarm Outputs Off
EtN2O Alarm switch On
High limit 55%
Low limit 0%
Priority Med
Alarm Outputs Off
FiN2O Alarm switch On
High limit 53%
Low limit 0%
Priority Med
Alarm Outputs Off
EtAA/FiAA Alarm switch On
High limit 30%
Low limit 0.0%
Priority Med
Alarm Outputs Off