10 - 17
10.7.5.3 Arrhythmia Alarm Shielding Rules
The following table explains how auidble and visual alarm indicate during arrhythmia alarm shielding period.
10.8 ST Segment Monitoring
ST monitoring is intended for adult, pediatric and neonatal patients.
10.8.1 ST Safety Information
• ST values may be affected by such factors as some drugs or metabolic and conduction disturbances.
• ST deviation is often calculated at a fixed offset from the J point. Changes in heart rate may affect ST.
• The ST deviation measurement algorithm has been tested for accuracy. The significance of ST 
segment changes needs to be determined by a physician.
• This monitor provides ST deviation level change information. The clinical significance of the ST level 
change information should be determined by a physician.
10.8.2 Enabling ST Monitoring
The ST monitoring function is disabled by default. Before you start ST monitoring, enable the ST function. To do 
so, follow this procedure:
1. Select the ECG numeric area or waveform area to enter the ECG menu.
2. Select the ST tab→ select the Setup tab.
3. Switch on ST Analysis.
Reliable ST monitoring cannot be ensured under the following situations:
■ You are unable to get a lead that is not noisy.
■ Arrhythmias, such as atrial fib or flutter, cause irregular baseline.
■ The patient is continuously ventricularly paced.
■ The patient has left bundle branch block.
In these cases, you may consider switching off ST monitoring.
Previous alarm Current alarm Alarm indication
Alarm in high priority 
chain
Alarm in high priority chain Alarm light and alarm tone
Alarm in medium priority chain  During the shielding period, alarm light and alarm 
tone are disabled. When the shielding period is 
reached, alarm light and alarm tone are 
reactivated.
Alarm in medium 
priority chain 
Alarm in high priority chain Alarm light and alarm tone
Alarm in the same medium priority 
chain, but with higher priority 
Alarm light and alarm tone
The same alarm reoccurs During the shielding period, alarm light and alarm 
tone are disabled. When the shielding period is 
reached, alarm light and alarm tone are 
reactivated. 
Alarm in the same medium priority 
chain, but with lower priority
During the shielding period, alarm light and alarm 
tone are disabled. When the shielding period is 
reached, alarm light and alarm tone are 
reactivated.
Alarm in other medium priority chain Alarm light and alarm tone