A - 25
A.13.11 BIS Specifications
EtCO
2
High (low limit + 2) to 99 mmHg 1 mmHg
EtCO
2
Low 1 to (high limit - 2)mmHg
FiCO
2
High 1 to 99 mmHg
EtO
2
High (low limit + 2%) to 100% 1%
EtO
2
Low 0% to (high limit - 2)%
FiO
2
High (low limit + 2%) to 100%
FiO
2
Low 18% to (high limit - 2)%
EtN
2
O High (low limit + 2) to 100% 1%
EtN
2
O Low 0 to (high limit - 2)%
FiN
2
O High (low limit + 2) to 100%
FiN
2
O Low 0 to (high limit - 2)%
EtHal/Enf/Iso High (low limit + 0.2) to 5.0% 0.1%
EtHal/Enf/Iso Low 0 to (high limit - 0.2)%
FiHal/Enf/Iso High (low limit + 0.2) to 5.0%
FiHal/Enf/Iso Low 0 to (high limit - 0.2)%
EtSev High (low limit + 0.2) to 8.0% 0.1%
EtSev Low 0 to (high limit - 0.2)%
FiSev High (low limit + 0.2) to 8.0%
FiSev Low 0 to (high limit - 0.2)%
EtDes High (low limit + 0.2) to 18.0% 0.1%
EtDes Low 0 to (high limit - 0.2)%
FiDes High (low limit + 0.2) to 18.0%
FiDes Low 0 to (high limit - 0.2)%
Standard Meet the standard of IEC 60601-2-26
Technique Bispectral index
Measured parameters EEG
BIS, BIS L, BIS R: 0 to 100
Calculated parameters SQI, SQI L, SQI R: 0 to 100%
EMG, EMG L, EMG R: 0 to 100 dB
SR, SR L, SR R: 0 to 100%
SEF, SEF L, SEF R: 0.5 to 30.0 Hz
TP, TP L, TP R: 40 to 100 dB
BC, BC L, BC R: 0 to 30
sBIS L, sBIS R: 0 to 10.0
sEMG L, sEMG R: 0 to 10.0
ASYM: 0 to 100%
Sweep speed 6.25 mm/s, 12.5 mm/s, 25 mm/s or 50 mm/s, ±10% error
Input impedance >5 MΩ
Noise (RTI) <0.3 μV (0.25 to 50 Hz)
Input signal range ±1 mV
EEG bandwidth 0.25 to 100 Hz