9 - 10
To exit the audio off status, press the Audio Pause quick key again.
WARNING
Pausing or switching off alarm sound may result in a hazard to the patient.
9.8 Resetting Alarms
Pressing the Alarm Reset quick key to reset the alarm system. When the alarm system is reset, the alarm reset
symbol displays in the system status information area for alarm symbols.
NOTE
If a new alarm is triggered after the alarm system is reset, the alarm reset icon will disappear and the
alarm light and alarm tone will be reactivated.
9.8.1 Resetting Physiological Alarms
Physiological alarms give different alarm indicators when the alarm system is reset:
The alarm sound is silenced.
A √ appears before the alarm message.
The color of the parameter numeric background corresponds with the alarm priority, but the parameter
numeric does not flash.
9.8.2 Resetting Technical Alarms
Technical alarms give different alarm indicators when the alarm system is reset:
Some technical alarms are cleared. The monitor gives no alarm indications.
Some technical alarms are changed to the prompt messages.
For some technical alarms, the alarm is silenced and a √ appears before the alarm message.
For details about the indications of technical alarms when the alarm system is reset, see D.2Technical Alarm
Messages.
9.9 Latching Alarms
The latching setting for physiological alarms defines how alarm indicators behave if you do not reset the alarms.
If you do not “latch” physiological alarms, their alarm indications disappear when the alarm condition ends.
If you “latch” physiological alarms, all visual and audible alarm indications remain until you reset the alarms.
For latched alarms the time when the alarm is last triggered is displayed behind the alarm message.
You can separately latch visual indications or simultaneously latch the visual and the audible indications.
When visual indications are latched, visual indications, including alarm lamp, alarm message and its
background remain when the alarm condition ends and the time when the alarm last triggered is displayed
behind the alarm message.
When audible indications are latched, the monitor issues alarm sounds when the alarm condition ends.
The alarm latch settings is password protected. For more information, see 27.4.3The Latching Tab.
NOTE
Changing alarm priority may affect the latching status of corresponding alarm. Determine if you
need to reset the alarm latching status if you changed the alarm priority.
When the alarm system is reset, latched physiological alarms are cleared.