4 - 5
4.4.9 Aldrete Score
Select Aldrete Score to show the latest score and scoring time. To understand the current patient status, select a
score for each item and then select OK to get a new score.
WARNING
• The Aldrete score and recommendation is for reference only. Clinicians must make the decision of
discharging the patient from recovery according to the patient’s actual condition.
4.5 The OxyCRG Screen
The monitor displays the OxyCRG screen by default when the neonatology department is selected. The OxyCRG
screen is available in any department setting, but only when Patient Category is set to Neo. This screen displays
6-minute HR/btbHR and SpO
2
trends, CO
2
/Resp compressed waveform, ABD parameters, and the latest ABD
events.
The OxyCRG function is intended for neonatal patients only.
4.5.1 Entering the OxyCRG Screen
To enter the OxyCRG screen, choose any of the following ways:
â–  Swipe left or right on the touchscreen with two fingers to switch to the OxyCRG screen.
â–  Select the OxyCRG quick key.
■ Select the Screen Setup quick key → select the Choose Screen tab → select OxyCRG.
■ Select the Main Menu quick key → from the Display column select Choose Screen → select OxyCRG.
4.5.2 OxyCRG Events
The following table lists the ABD events and their criteria:
NOTE
• The monitor records all ABD events for OxyCRG review, but only red events displays in the ABD list of
the OxyCRG screen.
4.5.3 The Display of the ABD Event Area
The ABD event area displays parameter values of currently active OxyCRG events and lists the latest red ABD
events.
Event type Description Remarks
A Apnea event: the apnea duration exceeds the threshold.
• A20: the apnea duration is greater or equal to 20 seconds.
• A15: the apnea duration is between 15 to 20 seconds (excluding 20 seconds).
• A10: the apnea duration is between 10 to 15 seconds (excluding 15 seconds).
A20 is a red event
B Bradycardia event: the duration of low heart rate, bradycardia, extreme bradycardia,
or asystole exceeds the threshold.
/
D Low SpO2 event: the SpO
2
value is lower than the SpO
2
Desat limit. /
BD Bradycardia and low SpO2 happen at the same time. /
AB Apnea and bradycardia happens at the same time. Red event
AD Apnea and low SpO2 happen at the same time. Red event
ABD Apnea, bradycardia, and low SpO2 happen at the same time. Red event