6 - 2
6.4 Setting Default Configuration
The monitor will load the pre-set default configuration in the following cases:
â–  A patient is admitted.
â–  A patient is discharged.
â–  Patient data is cleared.
â–  Patient category is changed.
To set the default configuration, follow this procedure:
1. Select the Main Menu quick key → from the Configuration column select Manage → input the required
password → select .
2. Select Select Default Config.
3. Select Load the Latest Config or Load Specified Config.
â—† When you select Load Specified Config, the restored configuration is subject to the patient category
(adult, pediatric or neonate). This configuration can be either factory configuration or a saved user
configuration. As an example, select Default Ped Config and then select Factory Default or user
configuration(s).
â—† When you select Load the Latest Config, the latest configuration is loaded when the monitor is
started or a patient is admitted.
6.5 Saving Current Settings
Current settings can be saved as a user configuration. Up to 25 user configurations can be saved.
To save current settings, follow this procedure:
1. Select the Main Menu quick key → from the Configuration column select Manage → input the required
password → select .
2. Select Save Current Settings.
3. Input the configuration name.
4. Select OK to save current settings as a user configuration.
6.6 Deleting a Configuration
To delete a configuration, follow this procedure:
1. Select the Main Menu quick key → from the Configuration column select Manage → input the required
password → select .
2. Select Delete Configuration.
3. Select the configuration you want to delete:
â—† In the Delete Configuration menu, selecting Local tab shows the existing user configurations on the
monitor.
â—† In the Delete Configuration menu, selecting USB Drive tab shows the existing user configurations
on the USB drive.
4. Select Delete.
5. Select OK.