C - 4
C.1.1.3 ST Default Settings
C.1.1.4 QT Default Settings
Item Default Setting
ST Alarm Mode Absolute
ST-I, ST-II, ST-III, ST-aVR, ST-aVL, ST-aVF, ST-V1, ST-
V2, ST-V3, ST-V4, ST-V5,ST-V6, ST-Va, ST-Vb (ST
Alarm Mode set to
Absolute)
Alarm switch Off
High limit 0.2 mV
Low limit -0.2 mV
Priority Med
Alarm Outputs Off
ST Single,
ST Dual (ST Alarm Mode set to
Relative)
Alarm switch Off
High limit 0.1 mV
Low limit -0.1 mV
Priority Med
Alarm Outputs Off
ST Analysis Off
ST Segment Auto
Show Markers Off
ST Point J+60 ms
Auto Adjust On
J48
ISO -80
Item Default Setting
QTc Alarm switch Off
High limit Adult: 500
Pediatric: 480
Neonate: 460
Priority Med
Alarm Outputs Off
QTc Alarm switch Off
High limit 60
Priority Med
Alarm Outputs Off
QT Analysis Off
QT Leads All