EasyManuals Logo
Home>Mindray>Medical Equipment>ePM 12MC

Mindray ePM 12MC User Manual

Mindray ePM 12MC
364 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #314 background imageLoading...
Page #314 background image
C - 4
C.1.1.3 ST Default Settings
C.1.1.4 QT Default Settings
Item Default Setting
ST Alarm Mode Absolute
ST-I, ST-II, ST-III, ST-aVR, ST-aVL, ST-aVF, ST-V1, ST-
V2, ST-V3, ST-V4, ST-V5,ST-V6, ST-Va, ST-Vb (ST
Alarm Mode set to
Absolute)
Alarm switch Off
High limit 0.2 mV
Low limit -0.2 mV
Priority Med
Alarm Outputs Off
ST Single,
ST Dual (ST Alarm Mode set to
Relative)
Alarm switch Off
High limit 0.1 mV
Low limit -0.1 mV
Priority Med
Alarm Outputs Off
ST Analysis Off
ST Segment Auto
Show Markers Off
ST Point J+60 ms
Auto Adjust On
J48
ISO -80
Item Default Setting
QTc Alarm switch Off
High limit Adult: 500
Pediatric: 480
Neonate: 460
Priority Med
Alarm Outputs Off
QTc Alarm switch Off
High limit 60
Priority Med
Alarm Outputs Off
QT Analysis Off
QT Leads All

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray ePM 12MC and is the answer not in the manual?

Mindray ePM 12MC Specifications

General IconGeneral
BrandMindray
ModelePM 12MC
CategoryMedical Equipment
LanguageEnglish

Related product manuals