Arrhythmia Threshold Default Settings
Pacer Not Pacing Off Prompt Off
Missed Beat Off Prompt Off
Nonsus V-Tach
CCU: On
Other departments: Off
Med Off
Vent Rhythm
CCU: On
Other departments: Off
Med Off
Pause Off Low Off
Irr Rhythm Off Prompt Off
A-Fib Off Prompt Off
PVCs/min
CCU: On
Other departments: Off
Med Off
Pauses/min
CCU: On
Other departments: Off
Med Off
Item Adult Pediatric Neonate
Asystole Delay 5 sec 5 sec 5 sec
Tachy 120 bpm 160 bpm 200 bpm
Brady 50 bpm 75 bpm 100 bpm
Extreme Tachy 160 bpm 180 bpm 220 bpm
Extreme Brady 35 bpm 50 bpm 60 bpm
Multif PVCs Window 15 beats 15 beats 15 beats
PVCs/min 10 10 10
Pause/min 8 8 8
Pause Threshold 2.0 sec 2.0 sec 2.0 sec
AF/Irr Rhy End Time 2 min 2 min 2 min
V-Tach Rate 130 bpm 130 bpm 160 bpm
V-Brady Rate 40 bpm 40 bpm 40 bpm
V-Tach PVCs 666
V-Brady PVCs 5 5 5
Item Alarm Switch Priority Alarm Outputs