27 - 3
27.3.4 The Location Tab
27.3.5 The Display Tab
27.4 The Alarm Settings
27.4.1 The Audio Tab
• The alarm volume escalation function is not applied to the latched alarms.
• The monitor provides the same alarm tone pattern for the remote device alarms as those for your 
monitor alarms.
Menu Item Default Setting Description
Location 1 - Location 10 / Selects where the patient goes after patient 
monitoring stops.
Menu Item Default Setting Description
Primary Screen Display Full Name On Selects whether patient name is displayed in the 
patient information area on the primary display.
Remote View Display Full Name On Selects whether patient name is displayed in the 
patient information area on the remote monitors 
when this monitor is viewed by other monitors.
Remote View Bedlist Display Full Name On Defines whether patient name is displayed in 
beds list on the remote monitors when this 
monitor is viewed by other monitors.
Menu Item Default Setting Description
Minimum Alarm Volume 2 /
Alarm Sound ISO Defines the alarm tone pattern to distinguish the 
heart beat tone, pulse tone, and keystroke tone 
by frequency.
High Alarm Interval 10 sec Defines the interval between alarm tones for the 
ISO mode.
Med Alarm Interval 20 sec
Low Alarm Interval 20 sec
Auto Increase Volume 2 Steps •
2 Steps: if an alarm is not reset within the 
designated delay time after the alarm occurs, 
the alarm volume automatically increases by 
two levels.
•
1 Step: if an alarm is not reset within the 
designated delay time after the alarm occurs, 
the alarm volume automatically increases by 
one level.
•
Off: if an alarm is not reset within the 
designated delay time after the alarm occurs, 
the volume of the alarm tone does not 
change.
Increase Volume Delay 20 sec Defines the delay time of alarm volume 
escalation