27 - 8
27.5.2 The GCS Tab
27.6 The Module Settings
27.6.1 The ECG Tab
27.6.2 The CO
2
 Tab
Menu Item Default Setting Description
Mild High limit 15 Selects the threshold and color of each 
consciousness level.
Low limit 13
Color White
Moderate High limit 12
Low limit 9
Color Yellow
Severe High limit 8
Low limit 3
Color Red
Menu Item Default Setting Description
ECG Standard AHA Selects the ECG standard according to the 
leadwires you are using.
QTc Formula Hodges Selects the QTc formula used to correct the QT 
interval for heart rate.
• Hodges:
 
• Bazett: 
 
•
•Fridericia: 
 
• Framingham: 
 
12-Lead Order No Selects whether to send the order of 12-lead 
interpretation report to the hospital information 
system while saving the report.
QTc QT 1.75 HearRate 60–()×+=
QTc QT
HearRate
60
--------------------------
1
2
---
×=
QTc QT
HeartRate
60
-----------------------------
1
3
---
×=
QTc QT 154 1
60
HeartRate
-----------------------------–
×+=
Menu Item Default Setting Description
Zero Recovery for 30s On After the zero calibration is completed, the CO
2
 
module reacquires the CO
2
 readings. 
•
On: During the reacquisition period, “Zero 
Recovering
” is displayed in the CO
2
 numeric 
area.
•
Off: During the reacquisition period, “Zero 
Recovering
” is not displayed in the CO
2
 
numeric area.