27 - 8
27.5.2 The GCS Tab
27.6 The Module Settings
27.6.1 The ECG Tab
27.6.2 The CO
2
Tab
Menu Item Default Setting Description
Mild High limit 15 Selects the threshold and color of each
consciousness level.
Low limit 13
Color White
Moderate High limit 12
Low limit 9
Color Yellow
Severe High limit 8
Low limit 3
Color Red
Menu Item Default Setting Description
ECG Standard AHA Selects the ECG standard according to the
leadwires you are using.
QTc Formula Hodges Selects the QTc formula used to correct the QT
interval for heart rate.
Hodges:
Bazett:
•Fridericia:
Framingham:
12-Lead Order No Selects whether to send the order of 12-lead
interpretation report to the hospital information
system while saving the report.
QTc QT 1.75 HearRate 60()×+=
QTc QT
HearRate
60
--------------------------


1
2
---
×=
QTc QT
HeartRate
60
-----------------------------


1
3
---
×=
QTc QT 154 1
60
HeartRate
-----------------------------


×+=
Menu Item Default Setting Description
Zero Recovery for 30s On After the zero calibration is completed, the CO
2
module reacquires the CO
2
readings.
On: During the reacquisition period, “Zero
Recovering
is displayed in the CO
2
numeric
area.
Off: During the reacquisition period, “Zero
Recovering
is not displayed in the CO
2
numeric area.