C - 2
C.1.1.2 Arrhythmia Default Settings
Arrhythmia Alarm Default Settings
Notch Filter On
Lead Set Auto
D12L(for 6-lead only) Off
Smart Lead On
Baseline Drift Removal (for 12-lead only) On
Waveform Layout (for 12-lead only) Standard
CrozFusion On
Display CrozFusion Off
QRS Volume
General, OR: 2
Other department: 0
QRS Threshold 0.16 mV
Paced Adult: Unspecified
Pediatric/Neonate: No
Pacer Reject Off
Item Default Setting
Item Alarm Switch Priority Alarm Outputs
Asystole On High, unadjustable Off
V-Fib/V-Tach On High, unadjustable Off
V-Tach On High, unadjustable Off
Vent Brady On High, unadjustable Off
Extreme Tachy On High, unadjustable Off
Extreme Brady On High, unadjustable Off
R on T
CCU: On
Other departments: Off
Med Off
Run PVCs Off Low Off
Couplet Off Prompt Off
Multiform PVC Off Med Off
PVC Off Prompt Off
Bigeminy
CCU: On
Other departments: Off
Med Off
Trigeminy
CCU: On
Other departments: Off
Med Off
Tachy Off Med Off
Brady Off Med Off
Pacer Not Capture Off Prompt Off