C - 6
C.1.3 SpO
2
Default Settings
Speed 6.25 mm/s
Auto Threshold Detection On
Item Default Setting
Item Default Setting
SpO
2
Alarm switch On
High limit Adult: 100%
Pediatric: 100%
Neonate: 95%
Low limit 90%
Priority Med
Alarm Outputs Off
SpO
2
Desat Alarm switch On
Low limit 80%
Priority High
Alarm Outputs Off
Sat-Seconds (for Nellcor SpO
2
)Off
NIBP Simul Off
FastSAT (for Masimo SpO
2
)Off
Display SIQ (for Masimo SpO
2
)Off
Sensitivity (for Mindray SpO
2
)Medium
Sensitivity (for Masimo SpO
2
)APOD
Averaging (for Masimo SpO
2
)8s
Display PI (for Mindray SpO
2
, Masimo SpO
2
)Off
Speed 25 mm/s
PR Alarm switch On
High limit Adult: 120
Pediatric: 160
Neonate: 200
Low limit Adult: 50
Pediatric: 75
Neonate: 100
Priority Med
Alarm Outputs Off
Alarm Source Auto
PR Source Auto
QRS Volume
General, OR: 2
Other departments: 0
Display PR Off