A - 18
A.13.8 C.O. Specifications
Sys High
IBP ≤ 50: (low limit + 2) to 50
IBP > 50: (low limit + 5) to 355
IBP ≤ 50: 1
IBP > 50: 5
Mean High
Dia High
Sys Low
IBP ≤ 50: -49 to (high limit - 2)
IBP > 50: 50 to (high limit - 5)
Mean Low
Dia Low
Art-S Extreme High High limit < 50: (High limit+ 1) to 360
High limit ≥ 50: (High limit+ 5) to 360
IBP ≤ 50: 1
IBP > 50: 5
Art-M Extreme High
Art-D Extreme High
Art-S Extreme Low low limit ≤ 50: -50 to (low limit- 1)
low limit > 50: -50 to (low limit- 5)
Art-M Extreme Low
Art-D Extreme Low
Standard Meet the standard of ISO 80601-2-56
Measurement method Thermodilution method
Measurement range C.O.:
TB:
TI:
0.1 to 20 L/min
23 to 43 °C
0 to 27 °C
Resolution C.O.:
TB, TI:
0.1 L/min
0.1 °C
Accuracy C.O.:
TB, TI:
±5% or ±0.1 L /min, whichever is greater
±0.1 °C (without sensor)
TB Operating mode Direct mode
Minimum time for accurate TB
measurement
10 s
Repeatability C.O.: ±2% or ±0.1 L/min, whichever is greater
Alarm range TB: 23 to 43 °C
Recovery time <15 s (after defibrillation)
Alarm limit Range Step
TB High (low limit + 1) to 43 °C
(low limit + 2) to 109.4 °F
0.1 °C
0.1 °F
TB Low 23 to (high limit - 1) °C
73.4 to (high limit - 2) °F