EasyManuals Logo

Mindray ePM 12MC User Manual

Mindray ePM 12MC
364 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #298 background imageLoading...
Page #298 background image
A - 18
A.13.8 C.O. Specifications
Sys High
IBP ≤ 50: (low limit + 2) to 50
IBP > 50: (low limit + 5) to 355
IBP ≤ 50: 1
IBP > 50: 5
Mean High
Dia High
Sys Low
IBP ≤ 50: -49 to (high limit - 2)
IBP > 50: 50 to (high limit - 5)
Mean Low
Dia Low
Art-S Extreme High High limit < 50: (High limit+ 1) to 360
High limit ≥ 50: (High limit+ 5) to 360
IBP ≤ 50: 1
IBP > 50: 5
Art-M Extreme High
Art-D Extreme High
Art-S Extreme Low low limit ≤ 50: -50 to (low limit- 1)
low limit > 50: -50 to (low limit- 5)
Art-M Extreme Low
Art-D Extreme Low
Standard Meet the standard of ISO 80601-2-56
Measurement method Thermodilution method
Measurement range C.O.:
TB:
TI:
0.1 to 20 L/min
23 to 43 °C
0 to 27 °C
Resolution C.O.:
TB, TI:
0.1 L/min
0.1 °C
Accuracy C.O.:
TB, TI:
±5% or ±0.1 L /min, whichever is greater
±0.1 °C (without sensor)
TB Operating mode Direct mode
Minimum time for accurate TB
measurement
10 s
Repeatability C.O.: ±2% or ±0.1 L/min, whichever is greater
Alarm range TB: 23 to 43 °C
Recovery time <15 s (after defibrillation)
Alarm limit Range Step
TB High (low limit + 1) to 43 °C
(low limit + 2) to 109.4 °F
0.1 °C
0.1 °F
TB Low 23 to (high limit - 1) °C
73.4 to (high limit - 2) °F

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray ePM 12MC and is the answer not in the manual?

Mindray ePM 12MC Specifications

General IconGeneral
BrandMindray
ModelePM 12MC
CategoryMedical Equipment
LanguageEnglish

Related product manuals