EasyManuals Logo

Mindray ePM 12MC User Manual

Mindray ePM 12MC
364 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #320 background imageLoading...
Page #320 background image
C - 10
IBP-M Alarm switch On
High limit
â–  Art/Ao/UAP/BAP/FAP/LV/P1/P2 arterial pressure
Adult: 110 mmHg
P
ediatric: 90 mmHg
Neonate: 70 mmHg
â–  PA
Adult: 20 mmHg
P
ediatric, Neonate: 26 mmHg
â–  ICP/RAP/LAP/UVP/P3/P4 venous pressure
Adult: 10 mmHg
P
ediatric, Neonate: 4 mmHg
â–  CVP
Adult: 14 cmH
2
O
Pediatric, Neonate: 5 cmH
2
O
Low limit
â–  Art/Ao/UAP/BAP/FAP/LV/P1/P2 arterial pressure
Adult: 70 mmH
g
Pediatric: 50 mmHg
Neonate: 35 mmHg
â–  PA
Adult: 0 mmHg
Pediatric, Neonate: 12 mmHg
â–  ICP/RAP/LAP/UVP/P3/P4 venous pressure
Adult: 0 mmHg
Pediatric, Neonate: 0 mmHg
â–  CVP
Adult: 0 cmH
2
O
Pediatric, Neonate: 0 cmH
2
O
Priority Med
Alarm Outputs Off
Art-S Extreme Alarm switch Off
High limit Adult: 175 mmHg
Pediatric: 130 mmHg
Neonate: 95 mmHg
Low limit Adult: 75 mmHg
Pediatric: 60 mmHg
Neonate: 50 mmHg
Priority High
Alarm Outputs Off
Art-D Extreme Alarm switch Off
High limit Adult: 105 mmHg
Pediatric: 80 mmHg
Neonate: 65 mmHg
Low limit Adult: 35mmHg
Pediatric: 30 mmHg
Neonate: 15 mmHg
Priority High
Alarm Outputs Off
Item Default Setting

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray ePM 12MC and is the answer not in the manual?

Mindray ePM 12MC Specifications

General IconGeneral
BrandMindray
ModelePM 12MC
CategoryMedical Equipment
LanguageEnglish

Related product manuals