D - 12
XX Conflicts
(XX refers to the module 
label)
Prompt / The same type of corresponding module being 
used exceeds the supported number. Remove 
the conflict module. 
XX Measurement has 
been closed
(XX refers to the module 
label)
Prompt / The parameter module is disabled. Switch on the 
module if you want to use it. For more 
information, see 
3.11.1Switching On or Off a 
Parameter
.
The display setup for XXX 
is disabled.
(XX refers to the 
parameter label)
Prompt / The parameter of the newly inserted module is 
not displayed on the screen. Select a desired area 
to display the parameter numerics and 
waveforms. For more information, see 
27.11The 
Other Settings
.
The patient data storage 
space is nearly full. Please 
delete some discharged 
patients.
Med B Delete unnecessary earlier discharged patient.
Alarm message Default priority Indication on 
alarm reset
Cause and solution