7 - 3
7.5.1.2 About the Remote View
The following figure shows the Remote View window.
7.5.1.3 Adding a Bed
You need to add the desired remote devices, and then the alarms from these devices can be watched on your 
monitor. To add a remote device, follow this procedure:
1. Enter the Select Bed window. To do so, choose either of the following ways:
◆ In the Remote View window, select Select Bed. For more information, see 7.5.1.1Entering the Remote 
View Window for entering the Remote View window.
◆ Select the   icon at the alarm watch tile if the tile is configured to display on the main screen.
2. In the Select Bed window, select a desired department. All the beds under this department will be listed.
(1)
Alarm watch area
◆ Display all the monitored remote beds. 
◆ Each bed displays the room number, bed number, connection status and alarm status. The 
background color indicates the alarm status on the corresponding bed. 
Background Color Description
Green No alarm is occurring to the bed.
Red The remote device is disconnected or a high priority alarm is occurring. The high 
priority alarm currently is the highest alarm level on the bed. If the remote device is 
disconnected, the   icon is displayed.
Yellow The medium priority alarm is occurring. The medium priority alarm currently is the 
highest alarm level on the bed.
Cyan The low priority alarm is occurring. The low priority alarm currently is the highest 
alarm level on the bed.
Grey The bed is in the standby mode.
(2)
Main body
Display the patient’s information, alarm status and messages, waveforms, measurements, etc. of the 
selected bed. This bed is called main bed.
(3) Remote View setup button: select it to enters the Remote View setup menu.