Accessible via MENU - [Operating] - [System setup] - [Weighing points] - [Weighing
point x] - [Calib] - [Modify] - [Corner correction] .
WP-D
Max 200t d= 0.02t
0.000 t
Calibration @admin
Local gravity
Max
Scale interval
Dead load
Calibration weight
Corner correction
9.81379 m/s²
200.000 t
0.020 t
7.510 t
11.000 t
10000 d
1 d
Corner correction @admin
LC 1
LC 2
LC 3
LC 4
2.319 t
6.496 t
4.406 t
5.491 t
You are free to choose any desired order.
Corner correction @admin
LC 1
LC 2
LC 3
LC 4
1.334 t
1.497 t
2.215 t
2.466 t
1. Select [Corner correction].
2. Press the [Platform 1] soft key.
3. Place the calibration weight on a corner.
4. Position (here: LC 2) is highlighted.
5. Conrm position.
6. Remove the calibration weight.
7. Repeat steps 1 to 6 for the remaining load cells.
Process Controller Maxxis 5 PR 5900 4 System setup
Minebea Intec EN-186