Function number 113 SetTare
Function number 114 ResetTare
Function number 115 SetTest
Function number 116 ResetTest
Function number 117 ResetPwf
Function number 118 SetFixTare (for function number 86, see Chapter 8.2.5.9)
Function number 119 GetFixTare (for function number 87, see Chapter 8.2.5.9)
Function number 121 ResetError
8.2.6 Example: reading the gross weight
Input range (MISO)
Byte Value Description
0
1
2
3
4 08 Read the gross weight (for func-
tion number 8, see Chap-
ter 8.2.5.8)
5
6
7
Output range (MISO)
Byte Value Description
0 00 Gross weight - byte 0 (MSB)
1 00 Gross weight - byte 1
2 04 Gross weight - byte 2
3 D2 Gross weight - byte 3 (LSB)
4 08 Gross weight request detected.
5 Write
Active
Power
Fail
6 Cmd
Busy
Cmd
Error
Tare
Active
Cal Ac-
tive
Test
Active
7 OutOf
range
Stand-
still
Inside
ZSR
Center
Zero
Below
Zero
Over-
load
Above
Max
ADU
Error
In direct access, status bits are
independent of the write or read
request.
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
The gross value (hex:000004D2 <=> 1234) can be read from bytes 03.
Process Controller Maxxis 5 PR 5900 8 Fieldbus interface
Minebea Intec EN-356