- [Print + tidy up selection of records], the Alibi memory is printed and then
deleted from the device, see Chapter 6.5.4.
3.7 Showing the calibration data
The calibration data is displayed in the format entered/determined during calibration.
Accessible via MENU – [Operating] - [System setup] - [Weighing points] - [Weighing
point x] - [Calib].
WP-A
Max 3000g d= 0.1g
00
+
g
Calibration @admin
Max
Scale interval
Dead load at
Max at
calibrated at
Sensitivity
3000.0 g
0.1 g
mV/V
mV/V
mV/V
µV/d
0.040368
1.177688
0.987884
0.471075
30000 d
1 d
102.832 g
3000.000 g
2516.5 g
98.14 cnt/d
New Modify
The calibration data is displayed:
- Scale interval and maximum capacity (Max)
- Scale interval(s)
- Dead load in weight and mV/V
- Weight and mV/V for maximum capacity
- Calibration weight* and corresponding mV/V
- Number of internal counts and voltage per scale interval
* After input with mV/V, the maximum capacity and the mV/V value entered are
displayed.
3.8 Increased resolution (10-fold) of the weight value
In the menu [Operating]- [System setup]- [Weighing points]- [Weighing point x] , the
weight is displayed with increased resolution (10-fold) for ve seconds after a weighing
point is selected, using the following keys:
- Key
- Function key F1 and/or F2 (if the function key has been assigned in the menu
[Operating]- [System setup]- [Operating parameters]- [F1/F2 key] ).
If the parameter [W&M] is selected for the weighing point, the weight value is marked as
an invalid weight with the symbol. After 2…3 seconds the display returns to normal
resolution.
3.9 Functions via the website
If the device is connected to the network, it can be displayed e.g. in the "Windows"
operating system under [Network].
3 Operation and control Process Controller Maxxis 5 PR 5900
EN-37 Minebea Intec