[W&M]
Setting for operation in legal metrology.
Selection: [None], OIML, NTEP (for USA) or NSC (for Australia), see Chapter 4.7.1.1.1.
[Max] (maximum capacity)
see Chapter 4.7.1.2.1.1.
[Scale interval]
See Chapter 4.7.1.2.1.2.
[Min. weight]
Minimum weight at which a print command can be triggered.
Setting range: 0…9999999 d.
For use in legal metrology a value of at least 20 d must be set.
[Update time]
Timeframe in which a new weight value is displayed.
Setting range: 0.1…2.0 s.
[Default] soft key
Settings are reset to factory settings.
4.7.7 Pendeo Truck
In this menu item (under Weighing points), the digital weighbridge load cell Pendeo
®
Truck is assigned to a logical weighing point and congured.
- [Interface] (serial), see Chapter 4.7.7.1.
- [Parameters] (load cells), see Chapter 4.7.7.2.
- [Calib], see Chapter 4.7.7.3.
- [Units] (display accuracy), see Chapter 4.7.7.4.
The digital load cells have been calibrated at the factory based on the acceleration of
gravity at Hamburg (9.81379 m/s
2
). The calibration data in the load cells are invariable.
The calibration data for the gravity acceleration at the place of installation can be adapted
only in the device and protected against overwriting (see Chapter 2.7). With applications
for use in legal metrology, the legal requirements and the conditions given on the test/
approval certicate must be taken into account when selecting the settings.
Accessible via MENU - [Operating] - [System setup] - [Weighing points] - [Weighing
point x] - [Weighing point] - [Pendeo Truck] .
3. Press the [Save] soft key to save the settings.
4 System setup Process Controller Maxxis 5 PR 5900