Setting range: 0…9999999 d
For use in legal metrology a value of max. 9 d = e must be set.
[Min. weight]
Minimum weight at which a print command can be triggered.
Setting range: 0…9999999 d
For use in legal metrology a value of at least 20 d must be set.
[Range mode]
Selection: single range, multi-range, multi-interval
For range selection for scales and settings, refer to Chapter 4.7.1.1.2 and 4.7.1.1.3
[Default] soft key
Settings are reset to factory settings.
4.7.1.1.1 W&M (use in legal metrology)
In this menu item (under Parameters), the parameters of the Internal weighing point
weighing electronics are congured for use in legal metrology.
Accessible via MENU - [Operating] - [System setup] - [Weighing points] - [Weighing
point x] - [Parameters] - [W&M] .
The following parameters can be selected: [None], OIML, NTEP (for USA) or NSC (for
Australia)
[None] [OIML] [NTEP] [NSC]
Gross weight display B B G G
0.2 mV/V @
30,000 d
0.2 mV/V @
3000 e
0.2 mV/V @
3000 e
0.2 mV/V @
3000 e
Recommended
min. measurement sig-
nals
0.4 mV/V @
60,000 d
0.4 mV/V @
6000 e
0.4 mV/V @
6000 e
0.4 mV/V @
6000 e
If legal-for-trade mode is switched on, the parameter settings (zerotrack etc.) must be
selected accordingly. The device does not perform a check of this.
A unique check number is created every time a calibration or changed parameters are
saved. This can be viewed in the menu [Operating] - [System information] - [Show
calibration check numbers] (see Chapter 5.7) and may also be written on the W&M label.
See also Chapter 4.7.1.2.
4.7.1.1.2 Range mode: Multi-range
In this menu item (under Parameters), a multi-range scale (of Class III or single-range
scale of Class I and II with variable scale interval) of the Internal weighing point weighing
electronics is congured
The multi-range scale is a scale with two or more weighing ranges with dierent
maximum capacities and scale intervals. There is only one load receptor, with each range
covering zero to its maximum capacity.
When [Range mode] = [Multi-range], the scale has up to three ranges with dierent
resolution.
3. Press the [Save] soft key to save the settings.
Process Controller Maxxis 5 PR 5900 4 System setup
Minebea Intec EN-98