[Zeroset range]
Determine a ±range around the zero point determined by the dead load during
calibration; within this range
- the displayed gross weight can be set to zero by pressing the zero-setting key (or
by a corresponding external command), and
- automatic zero tracking is active.
Setting range: 0.00…10000.00 d
For use in legal metrology a value ≤2% of the max must be entered, example: 60 d for
3000 e of Class III.
[Zerotrack]
The zero display is automatically maintained within set limits.
Selection: No/Yes
When [No] is selected, the next three parameters are not shown.
When [Yes] is selected, values for the next three parameters must be entered.
[Zerotrack range]
Range within which automatic zerotrack compensates for deviations.
Setting range: 0.25…10000.00 d
For use in legal metrology a value of 0.25 d must be entered.
WP-A
Max 3000g d= 0.1g
00
+
g
Weighing points @admin
1
2
3
Zerotrack step
Zerotrack time
Overload
Minimum weight
Range mode
0.25 d
1.0 s
9 d
20 d
single range
Default Save
[Zerotrack step]
If a weight change exceeds the adjusted value, automatic tracking does not function
any more.
Setting range for automatic tracking increments: 0.25…10 d
For use in legal metrology a value of 0.25 d must be entered.
[Zerotrack time]
Time interval for automatic zerotrack.
Setting range: 0.1…25 s
For use in legal metrology a value of 1 s must be entered.
[Overload]
Weighing range above the maximum capacity (Max) without error message.
4 System setup Process Controller Maxxis 5 PR 5900
EN-97 Minebea Intec