[Cuto frequency]
The smaller the cuto frequency, the slower the measurement and the more stable
the measurement result.
The cuto frequency for the low pass lter can be selected. The permitted range is:
0.01…40.00 Hz.
The available options depend on the measurement time.
The menu line [Cuto frequency] only appears if the digital lter is switched on.
[External load cell supply]
The external supply voltage is selected.
Selection: below or equal 8 V, above 8 V
WP-A
Max 3000g d= 0.1g
00
+
g
Weighing points @admin
1
2
3
Standstill time
Standstill range
Tare timeout
Zeroset range
Zerotrack
Zerotrack range
0.50 s
1.0 d
25 s
50.00 d
Yes
0.25 d
Default Save
[Standstill time]
The parameters [Standstill time] and [Standstill range] are used to dene the
standstill of the scale (stable balance).
The [Standstill time] parameter is entered in seconds. The permitted range is:
0.00…2 s.
The time can be entered from 0.00 to 2.00 seconds, but always makes up at least one
measurement time.
[Standstill range]
As long as the weight uctuations remain within this range, the device is determined
to be stable.
The [Standstill range] parameter is entered in "d". The permitted range is:
0.01…10.00 d.
For use in legal metrology, ≤1 d must be selected.
[Tare timeout]
Timeout for a tare/zeroset command that cannot be executed (e.g. due to mechanical
instability of the scale, incorrect lter setting, resolution too high, standstill condition
too strict).
This parameter is given in seconds. The permitted range is: 0.0…[2.5]…25 s.
At 0.0 s taring is only carried out when the scale is already stable.
Process Controller Maxxis 5 PR 5900 4 System setup
Minebea Intec EN-96