[/ Scale counts]
WP-D
Max 1000kg d= 1kg
0
kg
Weighing points @admin
1
2
3
Weighing point D
/ Scale Counts
Overload
Minimum weight
Update time
SPM marker address
User scale
1 cnt
9 d
0 d
0.3 s
32
Default Save
Number of scale counts for the weight increase entered under [Scale Weight /].
Setting range: 1…65000.
[Overload]
Weighing range above the maximum capacity (Max) without error message.
Setting range: 0…9999999 d.
[Min. weight]
Minimum weight at which a print command can be triggered.
Setting range: 0…9999999 d.
[Update time]
Timeframe in which a new weight value is displayed.
Setting range: 0.1…2.0 s.
[SPM marker address]
Free address for a double integer, i.e., SPM variable for the weight value.
SPM = Scratch Pad Memory.
[Default] soft key
Settings are reset to factory settings.
4.7.4 A+B (scale)
In this menu item (under Weighing points), the weighing electronics of two scales are
assigned to a logical weighing point and congured.
- [Parameters] (weighing electronics), see Chapter 4.7.4.1.
The scale A+B is selected if e.g. two vehicle weighbridges are to be dened as one scale.
The scale A+B can only be assigned to [Weighing point C].
The scale A and the scale B must satisfy the following requirements:
- Same scale interval
- Same unit
3. Press the [Save] soft key to save the settings.
In the [Weighing points] menu, the weight can be displayed with an increased
resolution (10-fold); see Chapter 3.8.
4 System setup Process Controller Maxxis 5 PR 5900