Accessible via MENU - [Operating]- [System setup]- [Weighing points]- [Weighing
point x]- [Weighing point]- [A+B] .
Weighing point C @admin
Weighing point not assigned
Save
Weighing point C @admin
Weighing point not assigned
Save
Not assigned
A+B
Liquid counter
User scale
xBPI scale
SBI scale
Pendeo Truck
Pendeo Process
1. Set up weighing point A and calibrate.
2. Set up weighing point B and calibrate.
3. Mark weighing point C and conrm.
4. Conrm [Weighing point].
A selection window opens.
5. Select and conrm [A+B].
6. Press the [Save] soft key to save the settings.
WP-C
Max 6000kg d= 1kg
617
+
kg
Weighing points @admin
Weighing point A
Weighing point B
Weighing point C
Weighing point D
Internal weighing point
Internal weighing point
A+B
SBI scale
Parameters
The weighing electronics are now assigned to the weighing point.
Process Controller Maxxis 5 PR 5900 4 System setup