Error display Possible causes
Set Max failed.
not enough d
This message is displayed if the selected resolution is
too low, e.g., 5 kg.
Set Max failed.
too many d
This message is displayed if the selected resolution is
so high that less than 0.8 internal counts per scale in-
terval (d) are available.
Set Max failed.
Max does not correspond to a
multiple of the scale interval.
This message appears if the maximum capacity [Max]
of the scale range (weighing range) is not an integer
multiple of the scale interval (1 d).
Set Max failed.
The weight cannot be adapted
to the scale.
No matching of the weight units, e.g. subsequent ch-
anging from kg to mg for [Max].
Set Max failed.
not enough μV/d for veriable
scale
This message appears if the selected resolution is so
high that less than 0.8 µV/e are available when use in
legal metrology as per OIML/NSC has been selected.
9.7.2 Determining the scale interval
WP-A
Max 3000g d= 0.1g
00
+
g
Max
Scale interval
Dead load at
Max at
calibrated at
Sensitivity
3000.0 g
0.1 g
mV/V
mV/V
mV/V
µV/d
0.040368
1.177688
0.987884
0.471075
30000 d
1 d
102.832 g
3000.000 g
2516.5 g
98.14 cnt/d
Save
STOP
Set LC verification
failed.
Max must be a multiple
of the scale division.
Ok
This message appears if the maximum capacity [Max] of the scale range is not an integer
multiple of the scale interval.
9.7.3 Determining the dead load
WP-A
Max 3000g d= 0.1g
00
+
g
Max
Scale interval
Dead load at
Max at
calibrated at
Sensitivity
3000.0 g
0.1 g
mV/V
mV/V
mV/V
µV/d
0.040800
1.177688
0.987567
0.470919
30000 d
1 d
102.832 g
3000.000 g
2516.5 g
98.14 cnt/d
by load by mV/V Save
STOP
Set dead load failed.
> phys. Max
Ok
Example display for the following error messages.
Process Controller Maxxis 5 PR 5900 9 Error messages
Minebea Intec EN-368