For use in legal metrology a value of at least 20 d must be set.
[Update time]
Timeframe in which a new weight value is displayed.
Setting range: 0.1…2.0 s.
[Default] soft key
Settings are reset to factory settings.
4.7.11 SMA scale
In this menu item (under Weighing points), the SMA scale weighing electronics are
assigned to a logical weighing point and congured.
- [Interface] (serial), see Chapter 4.7.11.1.
- [Parameters] (weighing electronics), see Chapter 4.7.11.2.
PR 5900 can communicate with scales via the SMA protocol. The devices can be
connected via serial interfaces (internal or PR 5900/04) (see the PR 5900 installation
manual under [Device installation] - [Hardware construction] - [RS-485 interface
(internal)] and under [Device installation] - [Accessories] - [PR 5900/04 2x RS‐485
interface]). Communication is serial.
Accessible via MENU – [Operating] - [System setup] - [Weighing points] - [Weighing
point x] - [Weighing point] - [SMA scale].
Weighing point C @admin
Weighing point not assigned
Save
3. Press the [Save] soft key to save the settings.
WP-C
Max 1.2kg d= 0.0001kg
00000
kg
Weighing points @admin
Weighing point A
Weighing point B
Weighing point C
Weighing point D
Internal weighing point
Internal weighing point
Mettler-Scale
Pendeo Process
Parameters
The weight value of the Mettler-Scale is displayed.
4 System setup Process Controller Maxxis 5 PR 5900
EN-221 Minebea Intec