Error display Possible cause
Set SPAN failed.
Weight >Max
This message appears when an attempt is made to
calibrate the scale with a weight heavier than the
[Max].
9.8 Error status in "LAST_ERROR"
An error occurs in the LAST_ERROR variable with actions such as SET_ZERO and
SET_TARE.
Possible error numbers are:
Error no. Cause
0 "OK" ok
1 "no wpt" The dened weighing point is not available.
2 The weighing point is occupied.
3 "not tared" Can't get TARE or NET if not tared
4 Incorrect weight unit
5 "range error" Weight range error
6 Test is active, no weight values.
7 Calibration is active, no weight values.
8 Scale is not at a standstill.
9 "is reactor"
10 No weighing electronics have been assigned.
11 The scale is not in batching mode.
12 The weighing point is not reserved.
13 Taring is active.
14 The weighing point is not being tested.
15 Weight not permissible in legal metrology (<0 or >Max).
16 Weight error
17 The scale is not ready.
18 Taring below zero is not possible.
101 The scale has rejected the command as incorrect.
102 Time limit for sending a command to the scale exceeded.
Process Controller Maxxis 5 PR 5900 9 Error messages
Minebea Intec EN-370