When the device is powered up, the following appears:
Checking…
Booting…
Restore…
The device is booting up.
No signal Error message: if no load cells are connected, see also Chapter 9.1.
No values from
scale
Error message: if there is no communication with the xBPI scale (see
also Chapter 9.2).
Error message: unable to read weight values from the ADC (analog-di-
gital converter); see also Chapter 9.1.
Scale not ready Error message: if no load cells or no scale is connected (see also Chap-
ter 9.1).
The weight display appears.
WP-A
Max 3000g d= 0.1g
00
+
g
0g 3000g
2.4 Switching o the device
Switching o the device, see Chapter 6.8.
2.5 Device warm-up time
A warm-up time of 30 minutes for the device is required before calibration is started.
2.6 Operating via front-panel keys/PC keyboard
The following options are possible:
- Via keys on the front of the device, see Chapter 2.1.3.1.
- Via an external PC keyboard (USB connection), see Chapter 2.1.3.4.
2.7 Overwrite protection
2.7.1 CAL switch
Overwrite protection can be activated using the CAL switch to prevent unauthorized
access to or overwriting of the calibration data/parameters.
The device can have up to 4 CAL switches. CAL switches 1 and 2 are located on the main
board and are accessible via two holes in the housing.
Process Controller Maxxis 5 PR 5900 2 Commissioning
Minebea Intec EN-20