Display Error and possible cause Remote dis-
play
Scale not ready No scale ready for weighing:
- The device is in the warm-up phase.
- The device is in automatic taring mode.
- The device was switched on with the scale loaded.
Error <
Wrong serial number Serial number of scale does not match the
number set in the device.
Error ?
Read cong failed The saved conguration and calibration data cannot be read.
- Restore the data saved in a backup (in menu [Operating] -
[System maintenance] - [Restore] - [SD card] ) to the device
(see Chapter 6.2.1).
- First set the weighing electronics (e.g. [internal weighing
point]) to [not assigned], save, and then reassign.
Error A
9.3 Error messages for Pendeo load cells
Display Error and possible cause Remote dis-
play
Overload The measuring signal is higher than Max + (x d):
- Wrong setting.
- Too much weight on the scale.
Error 2
Value exceeds display No display of weight values:
- Too many digits have been set.
Error 4
Incompatible units Incompatible weight units:
- Incorrect calibration values e.g. due to incorrect calibration.
Error 8
No values from scale No communication with Pendeo load cell:
- Cable break.
- Internal scale error.
- Scale not connected to power.
Error 9
Wrong conguration Number of load cells does not match the
conguration.
Error 13
Scale not ready No scale ready for weighing:
- At least 1 load cell gives an error status or is
defective (no communication)
Error <
Wrong serial number Serial number of scale does not match the
number set in the device.
Error ?
Warmup 123 The load cells are in the warmup phase,
duration approx. 30 s.
Error @