WARNING!
Changing the allocation of the fluid counter
after xBPI scale to Weighing point C
clears all adjustment data
and parameters for weighing point C.
Cancel Continue
This message appears if this weighing point is assigned to a dierent scale.
9.7 Error messages during calibration
- Determining the maximum capacity (MAX), see Chapter 9.7.1.
- Determining the scale interval, see Chapter 9.7.2.
- Determining the dead load, see Chapter 9.7.3.
- Calibrating with weight, see Chapter 9.7.4.
9.7.1 Determine MAX (maximum capacity)
WP-A
Max 3000g d= 0.1g
00
+
g
Calibration @admin
1
2
3
Max
Scale interval
Dead load at
Max at
calibrated at
Sensitivity
3000.0 g
0.1 g
mV/V
mV/V
mV/V
µV/d
0.040368
1.177688
0.987884
0.471075
30000 d
1 d
102.832 g
3000.000 g
2516.5 g
98.14 cnt/d
Save
STOP
Set Max failed.
> phys. Max
Ok
Example display for the following error messages.
Error display Possible causes
Set Max failed.
> phys. Max.
The maximum capacity of the scale can be increased
retroactively.
This message appears when the measurement signal
for the given maximum capacity would exceed the per-
missible input voltage.
Set Max failed.
below calibration weight
The maximum capacity of the scale can be increased
retroactively.
When the capacity is reduced, however, a message ap-
pears when the new maximum capacity falls below the
calibration weight ("calibrated at").
9 Error messages Process Controller Maxxis 5 PR 5900