9 Error messages
9.1 Error messages measuring circuit
The internal weighing electronics can generate error messages, which are shown on the
weight display.
Display Error and possible cause Remote dis-
play
Arith. error Internal arithmetic overow:
- Incorrect calibration values e.g. due to incorrect calibration.
Error 1
Overload The measuring signal is higher than Max + (x d):
- Wrong setting.
- Too much weight on the scale.
Error 2
No signal Measuring input open:
- The measuring signal is higher than the permissible range of
36 mV.
- Measuring cable is interrupted (cable break detection).
- Other hardware defect.
Error 3
Value exceeds display No display of weight values:
- Example: 30,000 kg scale and weight unit toggle to mg.
- > 9999000 mg (= 9,999 kg) this error message appears.
Error 4
No sense voltage No sense voltage:
- Load cells not connected.
- Sense line or supply line is interrupted.
- Wrong polarity or sense voltage is low.
Error 6
Negative input Negative measuring signal (< -0.1mV/V):
- Wrong polarity of load cell signal.
- Wrong polarity of load cell supply voltage.
Error 7
Incompatible units Incompatible weight units:
- Incorrect calibration values e.g. due to incorrect calibration.
Error 8
No values from scale Internal weighing point:
The measuring signal is higher than the permissible range of
36 mV. Cannot read weight values from ADC (analog-digital con-
verter).
- Error in weighing electronics board.
- Defective load cell.
- Cable break.
Error 9
Test active No gross weight value:
- Test has not yet ended.
Error 10