WP-A
Max 3000g d= 0.1g
00
+
g
Weighing points @admin
Weighing point A
Weighing point B
Weighing point C
Weighing point D
Internal weighing point
Internal weighing point
xBPI scale
Pendeo Process
Parameters Calib Units
When a CAL switch is closed, a tool tip is displayed in the menu [Weighing points].
The data under the soft keys [Parameters], [Calibration], and [Units] is displayed only.
WP-A
Max 3000g d= 0.1g
00
+
g
Calibration (CAL switch is closed) @admin
Max
Scale interval
Dead load at
Max at
calibrated at
Sensitivity
3000.0 g
0.1 g
mV/V
mV/V
mV/V
µV/d
0.040800
1.177298
0.987557
0.470919
30000 d
1 d
103.966 g
3000.000 g
2516.5 g
98.11 cnt/d
Example of calibration data.
2.7.2 Software
Overwrite protection can be activated via software to prevent unauthorized access to or
overwriting of the calibration data/parameters.
Note:
A unique check number is created every time a calibration or changed parameters are
saved. This can be viewed in the menu [Operating] - [System information] - [Show
calibration check numbers] (see Chapter 5.7) and may also be written on the W&M label.
Accessible via MENU – [Operating] - [System setup] - [Weighing points] - [Weighing
point x] - [Parameters] - [Settings locked].