WP-A
Max 3000g d= 0.1g
00
+
g
Weighing points @admin
Zerotrack step
Zerotrack time
Overload
Minimum weight
Range mode
0.25 d
1.0 s
9 d
20 d
single range
Default Save
WP-A
Max 3000g d= 0.1g
00
+
g
Weighing points @admin
Zerotrack step
Zerotrack time
Overload
Minimum weight
Range mode
0.25 d
1.0 s
9 d
20 d
single range
Default Save
One area
multiple range
multi-interval
WP-A
Max 3000g d= 0.1g
00
+
g
Zerotrack step
Zerotrack time
Overload
Minimum weight
Range mode
Range limit 1
0.25 d
1.0 s
9 d
20 d
multi-interval
1500.0 g
Default Save
4.7.1.2 Calib
In this menu item, the Internal weighing point weighing electronics are calibrated.
- [New], see Chapter 4.7.1.2.1.
- [Max] (maximum capacity), see Chapter 4.7.1.2.1.1.
1. Select and conrm [Range mode].
A selection window opens.
2. Select and conrm [Multi-interval].
3. Congure interval range 1: enter 1500.0 g for [Range limit 1].
4. Congure interval range 2: enter 2900.0 g for [Range limit 2].
5. Press the [Save] soft key to save the settings.
4 System setup Process Controller Maxxis 5 PR 5900
EN-101 Minebea Intec