4.7.1.2.1 New
In this menu item, a new calibration of the Internal weighing point weighing electronics
is conducted and congured.
The descriptions and settings of [Max], [Scale interval], [Dead load at] and [Max at] are
described in the following chapters:
- [Max] (maximum capacity), see Chapter 4.7.1.2.1.1.
- [Scale interval], see Chapter 4.7.1.2.1.2.
- [Dead load at], see Chapter 4.7.1.2.1.3 and, if correction is necessary, see also
Chapter 4.7.1.2.3.
- [Max at], see Chapter 4.7.1.2.1.4.
Accessible via MENU - [Operating] - [System setup] - [Weighing points] - [Weighing
point x] - [Calib] - [New] .
In this case: Select [Weighing point A] with [Internal weighing point] and press the
[Calib] soft key.
WP-A
Max 3000g d= 0.1g
00
+
g
Calibration @admin
Max
Scale interval
Dead load at
Max at
calibrated at
Sensitivity
3000.0 g
0.1 g
mV/V
mV/V
mV/V
µV/d
0.040368
1.177688
0.987884
0.471075
30000 d
1 d
102.832 g
3000.000 g
2516.5 g
98.14 cnt/d
New Modify
- The [New] soft key starts a new calibration, see Chapter 4.7.1.2.1.
- The [Modify] soft key (only use for small changes!) changes an existing
calibration, see Chapter 4.7.1.2.2.
The data from the last calibration is displayed.