EasyManua.ls Logo

Minebea Intec PR5900 - Page 111

Default Icon
382 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
[Save] is deactivated for new calibrations. The next menu item [Dead load at] is
activated.
Possible error messages, see Chapter 9.7.2.
WP-A
Max 3000g d= 0.1g
00
+
g
Calibration @admin
1
2
3
Max
Scale interval
Dead load at
Max at
calibrated at
Sensitivity
3000.0 g
0.1 g
mV/V
mV/V
mV/V
µV/d
0.040368
1.177688
0.987884
0.471075
30000 d
1 d
102.832 g
3000.000 g
2516.5 g
98.14 cnt/d
Save
STOP
Set LC verification
failed.
Max must be a multiple
of the scale division.
Ok
This message appears if the maximum capacity [Max] of the scale range is not an integer
multiple of the scale interval.
Press [OK] to cancel the setting.
4.7.1.2.1.3 Dead load at
In this menu item (under New/Modify), the dead load (empty scale or vessel) is dened
when calibrating the Internal weighing point weighing electronics.
Note:
If a linearization was carried out (see Chapter 4.7.1.2.1.4.4), the following message
appears once the parameter has been selected: "Cannot be changed while linearization
is active."
Only deleting of the linearization points deactivates the linearization mode!
Accessible via MENU - [Operating] - [System setup] - [Weighing points] - [Weighing
point x] - [Calib] - [New/Modify] - [Dead load at] .
WP-A
Max 3000g d= 0.1g
00
+
g
Calibration @admin
1
2
3
Max
Scale interval
Dead load at
Max at
calibrated at
Sensitivity
3000.0 g
0.1 g
mV/V
mV/V
mV/V
µV/d
0.040800
1.177298
0.987557
0.470919
30000 d
1 d
103.966 g
3000.000 g
2516.5 g
98.11 cnt/d
by load by mV/V Save
The scale interval (d) is calculated, based on the maximum weight value.
The conrmation is displayed with "setting scale interval…"
3. Press the [Save] soft key to save the calibration.
4 System setup Process Controller Maxxis 5 PR 5900
EN-109 Minebea Intec

Table of Contents

Related product manuals